CINRG - Support



CS-APC-3 Error ZERO_CNT

Issue Resolution

Instrumentation

Background

When solvent verification limits are much higher than the actual cleanliness of the solvent, the system will accept count data that is within the high limits but not representative of the actual cleanliness of the solvent. When this happens the reported counts on all samples will be lower than they should be because the corrections applied for the solvent background counts will be too high. This is not as issue for samples with high particle counts (i.e. > 15/13/11) as the error is minimal and insignificant, but this is not the case with very clean samples (i.e. < 14/12/10).

Resolution

The solvent verification limit parameters (Micron setting parameters) should be set based on the cleanest samples expected to be analyzed on the system and the highest dilution ratio used to dilute the samples. The parameters in the particle counter must be set to ensure the method requirement for the solvent not contributing more than 12.5% of the counts in the diluted sample is met on all samples.

Service Media

CINRG-CS-APC-SUPPORT-ZERO-COUNTS

Posted by Bill Quesnel on 09.05.2025

CINRG Support - Adjusting Solvent Verification Settings to Eliminate Negative Counts.

Read More


For more information or a quotation on CINRG instrumentation Contact Us Now!