Issue Resolution
Background
The laser diode in the Klotz particle counting sensor will deteriorate over time. As a result the noise level will continually increase until the sensor is not longer able to accurately resolve 4µm counts from the background noise. The time it takes for this to happen is highly dependent on the type of samples that are being routinely analysed.
Resolution
There are three options availabe, 1) repair the Klotz sensor, 2) replace the sensor with a new Klotz unit, or 3) request a service exchange sensor. Prior to replacing the sensor the noise level should be checked first to confirm this is the issue.
Parts Required
| Part Number | Description | |
|---|---|---|
| CS-CINSTAN-SCF | ![]() |
CINStan Super Clean Fluid (SCF) |
| CS-LDS-4550-REP | ![]() |
Klotz Particle Count Sensor Repair |
| CS-LDS-4550 | ![]() |
Klotz Particle Count Sensor |
| CS-LDS-4550-SEU | ![]() |
Klotz Particle Count Sensor Service Exchange |
Service Media
CINRG-CS-APC-22M-SUPPORT-HI-NOISE
CINRG Support - Determining Background Noise and Sensor Replacement Options.
CINRG-APC-22M-CALIBRATION-PROCEDURE
On-Site Calibration Manual - For CS-APC-22M Particle Counter.