Issue resolution

Detailed background, resolution steps, parts, and service media for CS-APC-22M HI_NOISE.

Issue Resolution

Background

The laser diode in the Klotz particle counting sensor will deteriorate over time. As a result the noise level will continually increase until the sensor is not longer able to accurately resolve 4µm counts from the background noise. The time it takes for this to happen is highly dependent on the type of samples that are being routinely analysed.

Resolution

There are three options availabe, 1) repair the Klotz sensor, 2) replace the sensor with a new Klotz unit, or 3) request a service exchange sensor. Prior to replacing the sensor the noise level should be checked first to confirm this is the issue.

Parts Required

Part Number Description
CS-CINSTAN-SCF CINStan Super Clean Fluid (SCF) CINStan Super Clean Fluid (SCF)
CS-LDS-4550-REP Klotz Particle Count Sensor Repair Klotz Particle Count Sensor Repair
CS-LDS-4550 Klotz Particle Count Sensor Klotz Particle Count Sensor
CS-LDS-4550-SEU Klotz Particle Count Sensor Service Exchange Klotz Particle Count Sensor Service Exchange

Service Media

PDF

CINRG-CS-APC-22M-SUPPORT-HI-NOISE

Posted by Bill Quesnel · 09.04.2025

CINRG Support - Determining Background Noise and Sensor Replacement Options.

PDF

CINRG-APC-22M-CALIBRATION-PROCEDURE

Posted by Alistair Geach · 03.02.2023

On-Site Calibration Manual - For CS-APC-22M Particle Counter.