Support · CS-APC-22M

Issue resolution

Detailed background, resolution steps, parts, and service media for CS-APC-22M LOOSE_TRAY.

Issue Resolution

Background

It is possible that the sample tray repeatedly comes loose during operation. The software may return the following error: Error code: Zaber reply error: “1 get pos”->”@01 0 OK IDLE FS -570811”

Resolution

The four (4) screws that hold the metal disc to the glass may come loose during operation. Follow the procedure in the bulletin to properly re-tighten the sample tray, and to modify the software settings to prevent the reoccurrence of the issue.

Service Media

PDF

CINRG-CS-APC-22M-SUPPORT-LOOSE-TRAY

Posted by Kannan Viswanathan · 09.05.2025

CINRG Support - Fixing a Loose Sample Tray.