CS-APC-2
CS-APC-2 Support
CS-APC-2 Support & Tips
Description
Solvent selection considerations for ASTM D7647
Configuring and Trouble-shooting the Baumer Level Sensor
Understanding Cup Calibration for the CS-APC Instruments
CS-APC-2 Errors
Error/Issue
Description
Probable Cause
13
Syringe parameter out-of-range
Incorrect software setting
20
X-axis movement fault
Dirt and lubrication
35
Missing encoder
Opto-detector board failure
45
Syringe overload
Multiple causes, flow restriction, tight syringe plunger, opto sensor failure, etc.
ZERO_CNT
Very low or zero 4µm counts
Insufficiently clean instrument solvent/diluent
COM_PORT
Instrument software error "Cannot perform operation on a closed port".
instrument cannot communicate with the Baumer level sensor and/or the Gilson Liquid Handler (XYZ table).